Description
Digital Pirani Vacuum Gauge- HPRGE-1GH-SP
Accuracy @ 25 °C |
From 1000 to 1 mbar |
+/- 2.5% |
From 1 mbar to 0.001 |
+/- 15% |
Special Features:
- specially designed for large volume evacuation vacuum readouts.
- High Accuracy At Higher Pressures from 1000 mbar to 1 mbar
- Very useful for accurate measurement of vacuum from 1000 mbar to 1X10e-4 mbar
- Brass body measurement sensor with standard KF-10 / KF-16 optional KF-25 end fitting.
TECHNICAL SPECFICATIONS – Sensor
MODEL |
S-EPR-GH |
On board sensor Display range |
1000 mbar to 1X10e-4 mbar |
Display type ( on sensor) |
Exponential format display x.ye-z) |
Units of measurement |
Mbar |
Temp range |
+2 to 50 deg C |
End connection |
Standard NW10 (optional NW16/25 available) |
Output |
4 to 20 mA |
Digital interfacing |
RS485 ( standard RTU modbus protocol) |
Zero & span adjustment |
Through Push button (at sensor end) |
Operating voltage |
+15V,0, -15V DC |
Mounting orientation |
Vertical only |
Environmental conditions for operation |
Not recommended for high moisture levels & corrosive gases evacuation |